The GC-AED Pre-Flight Checklist is part of our GC-AED Training Course.  It is a list of instrumental conditions to check prior to initiating a run on the GC-AED.  The checklist is also available as a PDF document that can be retrieved from our download page.

Is the GC Setup and Ready?

  • Check / replace septum
  • Check / replace injection port liner
  • Check / replace syringe
  • Fill solvent wash vials above line

Is the AED Setup and Ready?

  • Cooling water reservoir is full
  • Blue fly-ash filter is not wet
  • AED cavity pressure is approx 1.5 psi
  • AED helium supply pressure is 30 psi
  • Reagent gas pressures are OK
  • Plasma is lit and "looks" right

Is the ChemStation Setup and Ready?

  • On-line version of ChemStation is running and "Alive"
  • AED spectrometer is not realigning
  • AED spectral plot shows spectrum
  • Spectrum at 181nm with O2 Reagent "looks" right
 